FEATURES
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
  • 1 ml sample volume
  • Range from 400 to 700 nm.
  • Direct Concentration facility
  • Highly stable and Accurate
  • Battery/Mains Operated colorimeters also available

Digital Photo Colorimeters – 312 & 313 are the finest colorimeters that combine convenience in operation with high precision & accuracy in analysis.

Filter Photo Colorimeters – 321 & 322 are also available. These analog colorimeters give results in the form of %Transmission (%T), Absorbance (Abs.)

High standard glass filters covering complete visible range of 400 to 700 nm are mounted on a rotating disc. Highly sensitive photosensor and use of latest IC technology makes the instruments highly rugged, accurate & economical in maintenance. A minimum sample volume of 1ml is required to carry out the analysis. Provision has been made to switch on the lamp source only when required, thereby increasing the life of the photocell. The instruments are extremely useful for quantitative estimation of several chemical substances including blood chemistries in chemical/clinical labs, pharmaceutical industries, medicine plants, water treatment plants, agriculture universities etc.

SPECIFICATIONS

Model 312/313 321/331
High Std. Glass
Filters
312 – 8 Filters, 313 – 5 Filters 321 – 8 Filters, 322 – 5 Filters
Minimum Volume 1 ml
Display 2 ½ digit red seven segment
LED Analog Meter
Range 400 to 700 nm
Output Abs: 0 to 1.99 %T: 0 to 100, Abs: 0. to 1.99
Resolution Abs: 0.01 %T: 1, Abs: 0.01
Accuracy 0.5 % FS ± 1 Digit 2 % FS ± 1 Digit
Detector Photocell/Diode
Filters 8-Filters: 400, 450, 490, 520, 540, 570, 620, 680 nm

5-Filters: 450, 490, 540, 620, 680 nm
Light Source 6.8V, 300 mA Tungsten Lamp
Power 230V ±10%, 50Hz AC
Dimensions 195 x 235 x 105 mm (Approx.) 245 x 245 x 110 mm (Approx.)
Weight 2 Kg. (Approx.) 3 Kg. (Approx.)
Accessories Quartz cuvettes of 10 mm pathlength: (A set of 2), Operation Manual and Dust Cover
Note : Specifications are subject to change due to continuous improvements.