FEATURES
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
  • Double Beam Optics
  • Scanning UV-VIS Spectrophotometer
  • 24 x 2 Line Alphanumeric Backlit LCD
  • Compact, Modular And Economical
  • Automatic Wavelength Calibration
  • Data Saving And Retrieval Facility
  • Programmable Lamp Change Over Wavelength
  • Microprocessor Based With Printer Interface
  • Optional PC Compatibility With RS-232C Interface
  • Automatic 6-Position Cuvette Holder For 10 mm Pathlength Cuvettes
  • Fully Automatic, User Friendly And Menu Driven Software
  • Self Check And Self Diagnostic Features

Double Beam UV-VIS Spectrophotometer is a compact and easy to operate instrument for spectrophotometric analysis of any concentration. The output is available on a 24 x 2 line alphanumeric backlit LCD display in terms of %Transmission (%T), Absorbance (Abs.), and Concentration (Conc.). The instrument has 1000 sample storage memory, printer output for printing data with any dot matrix printer. RS-232C interface for computer can be provided on request. The use of microcontroller had made it an extremely versatile and state-of-the-art instrument. Facility for using 10 mm pathlength cuvette is provided.

The instrument operates at wavelength between 190 to 1100 nm. The use of diffraction grating in the monochromator produces a linear spectrum. The use of grating having very high resolution of 1200 lines/mm ensures a narrow bandwidth. Deuterium lamp (D2) is used in the range 190 to 380 nm and Tungsten Halogen lamp (W) is used in the range 380 to 1100 nm.

SPECIFICATIONS

SPECTRAL
Range 190 to 1100 nm
Bandwidth 2 nm
Accuracy ± 0.5 nm
Readability 0.1 nm
Repeatability ± 0.2 nm
PHOTOMETRIC
Range %T: 0 to 100.0%, Abs: 0 to ± 3.000
Accuracy ± 0.005 Abs at 1.0 Abs, ± 0.010 Abs at 1.5 Abs
Readability %T: 0.1, Abs: 0.001
Repeatability ± 0.002 Abs at 1.5 Abs
Stray Light Less than 0.1%T at 260 nm with KI 10 g/l
System Double Beam Optics
Light Sources Deuterium Lamp (D2), Tungsten Halogen Lamp (W)
Monochromator Czerny-Turner type with 1200 lines/mm
Detector Wide range highly sensitive photodiode
OPERATIONAL
Baseline Correction Automatic baseline correction
Measuring Mode Scan (Transmission & Absorbance), Photometric (Multi & Serial),
Scan Speed Slow, Medium, & Fast
Scan Interval 5 to 99 seconds
Sample Holder Automatic 6-Position cuvette holder for 10 mm pathlength cuvettes
Filters Auto, in-built filter wheel
Data Storage 1000 samples storage memory
DATA PRESENTATION
Control Microcontroller based
Display 24 x 2 line alphanumeric backlit LCD display
Printer Interface Available for any dot matrix centronics printer
Power 230V ± 10% AC, 50Hz, 150VA
Dimensions 615 x 495 x 155 mm (L x B x H) (Approx.)
Weight 28 Kg (Approx.)
Accessories Quartz cuvettes of 10 mm pathlength: (A set of 2), Polystyrene Disposable Cuvettes 10 mm pathlength: (A set of 5), Main Lead, Operation Manual and Dust Cover
Note : Specifications are subject to change due to continuous improvements.